7 User menu
Parameter
Heart rate
Heart rate
Pulse rate
Pulse rate
SpO
2
saturation
SpO
2
Alarm
saturation
limit
settings
NIBP systolic
NIBP systolic
NIBP
diastolic
NIBP
diastolic
Autom.
alarm
limits
112
EN
MEDUCORE Standard
Possible
Description
values
35/min-250/min,
in increments of 5 Here you can set the
30/min-245/min,
upper ( ) and lower
in increments of 5
( ) limits as of which
the device must emit
35/min-250/min,
an alarm.
in increments of 5
According to the
30/min-245/min,
values set, the setting
in increments of 5
ranges of the upper
66 %-100 %
and lower limits
depend on each other
as follows:
65 %-99 %
The setting range for
45 mmHg-
the upper alarm limit
260 mmHg,
ends at least one
in increments of 5
setting value above
40 mmHg-
the set value for the
255 mmHg,
lower alarm limit. The
in increments of 5
setting range for the
lower alarm limit ends
25 mmHg-
at least one setting
200 mmHg,
value below the set
in increments of 5
value for the upper
20 mmHg-
alarm limit.
195 mmHg,
in increments of 5
Here you can set the
automatic alarm
limits. The device sets
the alarm limits for the
10 %
physiological alarms
20 %
automatically. The
30 %
deviation is 10 %,
20 % or 30 % from
the values measured
at the time of
activation.
2
Factory setting for patient
group
Adult
Child
120/min
150/min
50/min
50/min
120/min
150/min
50/min
50/min
100 %
100 %
85 %
85 %
220 mmHg 145 mmHg 100 mmHg
75 mmHg
75 mmHg
110 mmHg 100 mmHg 70 mmHg
35 mmHg
35 mmHg
20 %
Infant
200/min
100/min
200/min
100/min
95 %
85 %
50 mmHg
30 mmHg