15 Technical data
15.1 Device
Specification
Product class according to Directive
93/42/EEC
Dimensions (W x H x D)
Weight:
Without battery
With battery
Continuous operation:
•
Temperature range
•
Humidity
•
Air pressure
•
Height above sea level
Transient operation (temporary
operation at the site of use):
•
Temperature range
•
Humidity
•
Air pressure
•
Height above sea level
Storage/transport*:
•
Temperature range
•
Humidity
•
Air pressure
•
Height above sea level
Electrical connection (rated voltage)
Max. power consumption
Vehicle electrical system operation:
•
Rated voltage
•
Max. internal resistance of vehicle
electrical system
•
Maximum current consumption
Operating mode
The technical data of components and accessories may deviate.
You should also note the sections which follow, as well as the
Instructions for Use for the component/accessory part.
Device
IIb
242 mm x 137 mm x 130 mm
2.25 kg
2.75 kg
0 °C to +40 °C
15 % rh to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
-20 °C to +55 °C
15 % rh to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
-40 °C to +70 °C
15 % rh to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
12 V to 15.1 V
30 W
12 V
500 mΩ
< 3 A
Continuous operation
MEDUCORE Standard
15 Technical data
2
EN
175