8 Operator menu
Parameter
Disable functions
Export status log
to SD card
Export device
configuration to
SD card
SD card
Import device
configuration
from SD card
Format SD card
Service reminder
Factory settings
Software
140
EN
MEDUCORE Standard
Possible
Description
values
Here you can disable certain
functions and protect against
unauthorized access with a
code prompt:
Each function can be
protected individually and
also assigned an individual
code.
Upon delivery from the
factory, the code is 0000 and
the functions are not
protected.
Here you can export the
status log to the SD card.
Here you can export the
currently set values of the
operator menu to the SD card
so that they can be imported
by another device.
Here you can import the
settings which have been
copied to an SD card from the
operator menu of another
device
Here you can format the
inserted SD card. This deletes
all the data from the SD card
Activated
Here you can set whether the
device should output a
reminder when there are less
Deactivated
than 30 days until the next
service interval.
Here you can reset the device
to the factory settings.
Here you can update the
device software.
2
Manual mode
Curve view*
Venous stasis
Factory
setting
Activated
-
-

Hide quick links: