Operation; Switching The Device On - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

5 Operation

5
Operation

5.1 Switching the device on

Requirement
76
EN
MEDUCORE Standard
The ECG cable and the defibrillation electrodes are not
connected to the patient.
A fully charged battery is inserted in the device.
1. Briefly press the On/Off button
An automatic self-test starts, which runs through the following
sequence:
Alarm light flashes and test tone sounds
The start screen appears
Shock standby indicator is illuminated
The self-test is successful when all of the steps have been
completed.
When doing so, please note:
If the device was switched off for ≥ 30 s: The start menu
appears. The device starts with the presets in the operator
menu.
If the device was switched off for < 30 s and, if whilst
switched on previously, patient measurements were taken
or an event was saved manually: The device skips the start
menu and starts in the preset start mode and with the
preset start view. The settings in the user menu from the
last session are retained and the device assigns the session
data to the last session.
If the ECG cable and the defibrillation electrodes were
already connected to the patient at the start of the self-test:
The device skips the test of the ECG module and the
defibrillation module.
2. If one or more conditions are not met: Do not operate the
device.
3. Perform a function check
check", page
148).