Weinmann MEDUCORE Standard2 Instructions For Use Manual page 40

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

3 Description
Status indicator on
the battery
40
EN
MEDUCORE Standard
Status indicator on
the device display
2
Meaning
Battery status < 10%
On the display:
The last remaining segment in the battery status
symbol is red.
The message Battery weak appears in the
display.
The device outputs in AED mode: Battery weak.
Battery is deeply discharged. Charge battery in the
device for 24 hours. After 24 hours:
Green LED is lit: Battery fully charged and ready
for use
Red LED or no LED is lit: Battery defective. Replace
battery.
Battery is empty
Battery empty appears on the display and the
device outputs in AED mode:
Battery empty.
The device can still be used for approx. 15 minutes.
Battery is defective. Replace battery.
Battery is defective.
or
No battery.
or
Battery not at suitable temperature.
Green arrow: Battery is charging