Advertisement

5 Operation
Result
5.8.3 Venous stasis
96
EN
MEDUCORE Standard
5. Press the Start function button.
or
Press the NIBP button
An NIBP measurement starts. At the end of the measurement,
the device shows the systolic and diastolic arterial pressure, the
timer expires and the next measurement starts automatically.
6. Evaluate the NIBP measurement.
7. If necessary: Cancel the NIBP measurement with the Stop
function button.
The device releases the pressure from the NIBP cuff.
8. If necessary: Set alarm limits in the user menu
Alarm settings", page
or
Set automatic alarm limits in the application menu
Application menu", page
9. If necessary: Press the Back function button.
or
Press the NIBP button
The device ends the NIBP function mode and switches to the
set mode.
The interval measurement has been performed.
With venous stasis (iv function button), the device inflates the NIBP
cuff and maintains this pressure for the time preset in the operator
menu. The venous blood return flow is impeded and the user can
puncture one of the patient's veins. The venous stasis function is
only available to the Adult patient group.
If you are the operator of the device and have access to the
operator menu, you can disable the Venous stasis function:
Operator menu | System settings | Disable functions | Venous stasis
(see "8.8 System settings", page
2
for < 2 s.
111).
106).
for < 2 s.
137).
(see "7.3.1
(see "6