Preparation; Mounting The Device; Connecting To A Power Supply - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

4

Preparation

4.1 Mounting the device

4.2 Connecting to a power supply

The device is mounted on a portable system as standard and is
ready for use. Follow the Instructions for Use for the portable
systems.
Risk of injury due to missing battery!
During line operation, defibrillation is not possible without the
battery. Line operation without the battery impairs the operational
readiness of the device.
⇒ Only operate the device with the battery inserted.
Damage to the device caused by ingress of liquids!
The device is only protected from water jets as per IP55 when the
battery is inserted, the water jet protection for the SD card slot is
closed and the lines for the ECG, SpO
connecting tube including NIBP cuff are connected. Ingress of
liquids and dust may damage the device, components, and
accessories.
⇒ Do not immerse the device, components, or accessories in
liquids.
⇒ Only operate the device with the battery inserted.
⇒ Always close the water jet protection of the SD card slot.
⇒ Always leave the lines for ECG, SpO
connecting tube including NIBP cuff connected.
1. Check battery status
indicator", page
2. If necessary: Charge battery
in the device", page
, trunk cable and NIBP
2
2
(see "3.5 Battery and battery status
39).
(see "4.3.2 Charging the battery
55).
MEDUCORE Standard
4 Preparation
, trunk cable and NIBP
2
EN
53