Maintenance; General Instructions; Intervals - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

12 Maintenance

12 Maintenance

12.1 General instructions

12.2 Intervals

Part concerned
Device
Battery
Accessories
170
EN
MEDUCORE Standard
Maintenance, safety checks (Germany only), inspections, and
repairs must only be carried out by the manufacturer or a
technician specifically authorized by the manufacturer.
Interval
Maintenance-free
Annual safety check recommended Manufacturer or a technician
Metrological check every two years
Maintenance-free
When stored in the device: Charge every 3 months.
When not stored in the device: Charge every 5 months.
Recommendation: Replace battery after 2 years.
Please follow the Instructions for Use supplied with the accessories.
Should the accessories not come with their own Instructions for Use, the
same intervals as for the device shall apply.
2
Maintenance by
specifically authorized by the
manufacturer