15 Technical data

15.3 Battery

Specification
Type
Dimensions (W x H x D)
Weight
Shock capacity (at 20 °C with new and fully charged
battery)
Monitoring capacity
Battery run time for resuscitation
Rated capacity
Rated voltage
Charging time (0 % to 90 %)
Charging temperature
Service life

15.4 Power supply unit and charger

Specification
Operation:
Temperature range
Humidity
Air pressure
Height above sea level
Max. power consumption
Input voltage
Rated voltage output
Degree of protection against
Ingress of solid objects
Ingress of dust
Ingress of water with harmful effect
Disconnection from line power
178
EN
MEDUCORE Standard
Battery
Li-ion
97 mm x 127 mm x 33 mm
450 g
350 shocks at 200 J
Approx. 5 h
4 h with shocks of 200 J
4.2 Ah (≥ 46.4 Wh)
10.8 V
Approx. 3.5 h
0 °C to +45 °C
At least 300 cycles or 2 years
Power supply unit and charger
0 °C to +40 °C
5 % rh to 95 % rh, no condensation
700 hPa to 1100 hPa
-500 m to 3000 m
100 W
100 V-240 V~/50 Hz-60 Hz
15 V
IP21: Protected against foreign solids and dripping
water
Pulling out the power plug disconnects the device
from line power on all poles.