11.2.3 Low-priority alarm (turquoise)
Alarm
Battery operation
Insert battery
Date wrong
Internal memory full
SD card defective
Insert SD card
SD card full
SpO
sensor defective
2
Cause
Line power too low.
Disconnect the line power supply
by removing from the wall
mounting.
Power outage
Battery not inserted or incorrectly
inserted (monitor mode).
Date read-out error.
Internal module defective.
No SD card inserted and session
duration > 4 h-10 h.
SD card write/read error.
No SD card/SD not detected.
SD card full.
Pulse oximetry sensor connecting
cable defective.
Pulse oximetry sensor defective.
11 Alarms and faults
Remedy
Alarm goes off automatically after
10 s. Restore line power.
Insert battery correctly
53).
Insert battery
Reset date.
If this recurs: Have the device
repaired.
To continue to record the session:
Insert SD card
copy data to SD card, and resume
session with SD card.
Continue with session and do not
record any more data.
Insert new SD card
73), switch off device
77), and switch device back on
after at least 30 s
Continue with session and do not
record any more data.
Insert new SD card
73), switch off device
77), and switch device back on
after at least 30 s
Insert new SD card
73), switch off device
77), and switch device back on
after at least 30 s
Erase data on current SD card.
Replace pulse oximetry sensor
connecting cable.
Replace pulse oximetry sensor.
MEDUCORE Standard
(see 4.2, p.
(see 4.2, p.
53).
(see 4.8.1, p.
73),
(see 4.8.1, p.
(see 5.2, p.
(see 5.1, p.
76).
(see 4.8.1, p.
(see 5.2, p.
(see 5.1, p.
76).
(see 4.8.1, p.
(see 5.2, p.
(see 5.1, p.
76).
2
EN
163