Weinmann MEDUCORE Standard2 Instructions For Use Manual page 167

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Fault
Battery not charging although it is
not full
11.3.4 6-lead ECG
Fault
The Plug in ECG connector
alarm appears although the ECG
cable is connected to the device
The ECG cable is not connected to
the device but the Plug in ECG
connector alarm does not appear
The Check seating of ECG
electrodes alarm appears
although the ECG electrodes are
attached to the patient
ECG cable not attached or not
correctly attached to the patient,
but the alarm Check seating of
ECG electrodes does not appear
11.3.5 Pulse oximetry
Fault
The Plug in SpO
sensor alarm is
2
emitted even though the pulse
oximetry sensor is connected to
the device
Cause
Battery temperature < 0 °C or
> 45 °C.
Battery defective.
Cause
Device does not detect connected
ECG cable.
Device incorrectly detects ECG
cable as connected.
The alarm is only emitted if an ECG
has been successfully derived since
switching on.
ECG electrodes not correctly
attached to the patient.
Device incorrectly detects ECG
cable as attached to the patient.
The alarm is only emitted if an ECG
has been successfully derived since
switching on.
Cause
Device does not detect connected
pulse oximetry sensor.
11 Alarms and faults
Remedy
Charge battery within permitted
temperature range
178).
Replace battery.
Remedy
Check connection.
Replace ECG cable.
Have the device repaired.
Have the device repaired.
-
Attach ECG electrodes to the
patient correctly
Replace ECG cable.
Have the device repaired.
Have the device repaired.
-
Remedy
Check connection.
Replace pulse oximetry sensor
connecting cable.
Replace pulse oximetry sensor.
Have the device repaired.
MEDUCORE Standard
(see 15.3, p.
(see 4.6, p.
64).
2
EN
167