4 Preparation
70
EN
MEDUCORE Standard
When doing so, please note:
Select a
smaller one
The index printed on the NIBP cuff must be within the
printed range.
If the index marking does not extend into the printed
range: Select a larger NIBP cuff.
If the index marking extends beyond the printed range:
Select a smaller NIBP cuff.
In the case of disposable cuffs for neonates, use adapter
tube (WM 45467).
3. Switch on the device
76).
Risk of injury due to incorrectly selected patient group!
If the incorrect patient group is selected, the contact pressure of
the NIPB cuff may be too high and injure the patient.
⇒ Adapt the patient group to the patient.
⇒ If necessary: Change the patient group in the application
menu.
4. Select the patient group in the start menu.
2
Right size
RANGE
(see "5.1 Switching the device on", page
Select a
larger one