5 Operation
Result
82
EN
MEDUCORE Standard
If you operate the device via the line power and the inserted
battery is defective or if the battery does not have sufficient
capacity to charge the capacitor for shock energy in AED mode,
the device guides you through cardiopulmonary resuscitation
without creating a shock standby. If you subsequently insert an
undamaged and sufficiently charged battery, the device starts the
cardiac rhythm analysis immediately and prepares for a shock
standby.
The device performs a cardiac rhythm analysis. The cardiac rhythm
analysis has one of two results:
Shock required
or
Shock not required
Shock required
The device performs a cardiac rhythm analysis, charges for shock
delivery and outputs the message:
Voice prompt
Stand clear of the patient
Cardiac rhythm is being analyzed
If, based on the cardiac rhythm analysis, the device determines that
a shock is required, the device outputs the message:
Voice prompt
Shock required
Press shock button
The shock button flashes
2
(see " Shock required", page 82)
(see " Shock not required", page 84)
AED instruction
Stand clear of the patient
Analysis
AED instruction
Shock required
Press shock button
and an audio alarm sounds.