11 Alarms and faults
Alarm
Check seating of SpO
2
SpO
signal quality
2
Plug in SpO
sensor
2
164
EN
MEDUCORE Standard
Cause
Pulse oximetry sensor not attached
sensor
or not correctly attached to the
patient.
SpO
signal quality < 40 %.
2
SpO
connector of the pulse
2
oximetry sensor connecting cable
not connected or incorrectly
connected to the device.
2
Remedy
Attach pulse oximetry sensor to
the patient correctly
(see 4.5, p.
60).
Attach pulse oximetry sensor to
the patient correctly
(see 4.5, p.
60).
If medically indicated, perform
suitable treatment.
Connect SpO
connector of the
2
pulse oximetry sensor connecting
cable to the device correctly.