Weinmann MEDUCORE Standard2 Instructions For Use Manual page 189

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Delivered energy as a function of patient impedance*
Selected
Patient impedance in Ω
energy in J
25
0.4
1
2.6
5
5.2
10
7.8
15
10.2
20
15.4
30
19.8
40
26
50
38.4
75
53
100
63.6
120
79.6
150
128.9
200
50
75
100
1
2.5
2.4
5.4
7.8
6.8
10.8
13.2
11.6
15.7
20.4
18.4
19.9
25.6
23.4
30.2
42.1
40.2
39
51.4
49.8
50.2
72.1
72
73.8
106.6
107.7
102
135.6
134.1
122.3
153.7
148.3
152
181.7
170.9
201.2
211.6
208.3
*The selected energy differs from the delivered energy due to the
laws of physics. The delivered energy depends on the patient
impedance (see also IEC 60601-2-4:2010 Chapter 201.12.1 and
Annex AA).
125
150
175
2
1.8
1.7
6
5.4
4.9
10.2
9.3
8.4
16.5
15.3
13.6
21
19.2
17.5
36.7
33.9
31.1
46
42.3
38.8
65.7
60.3
55.6
100.7
92.9
86
123.2
113.3
104.9
137.2
127.1
116.8
157.7
146.3
134.7
205.6
196
187.5
MEDUCORE Standard
15 Technical data
Precision
of
delivered
200
energy
1.6
±3 J
4.4
±3 J
7.6
±3 J
12.4
±3 J
16
±15 %
28.8
±15 %
36.4
±15 %
51.6
±15 %
79.9
±15 %
96.3
±15 %
107.9
±15 %
126.7
±15 %
179.5
±15 %
2
EN
189