Weinmann MEDUCORE Standard2 Instructions For Use Manual page 121

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Parameter
Adult AED settings/child AED settings
CPR voice
prompts
Audio outputs
Metronome
Metronome
frequency
Start analysis
automatically
Pictograms in
parameter view
Charging during
analysis
Possible
Description
values
Here you can set whether the
Activated
voice prompts for guiding
chest compression (Ventilate
twice and Carry out
Deactivated
cardiopulmonary
resuscitation) should be
output.
Activated
Here you can activate or
deactivate the metronome
Deactivated
output.
100/min to
Here you can set the
120/min
frequency of the metronome.
Activated
Here you can set whether the
cardiac rhythm analysis starts
Deactivated
automatically, or manually at
the touch of a button.
Activated
Here you can set in the AED
mode whether, instead of
Deactivated
parameters, pictograms are
shown in the parameter view.
Here you can set whether the
defibrillation capacitor is to
Activated
be charged at the same time
as the cardiac rhythm
analysis and thus
Deactivated
independently of the analysis
result. If this setting is
deactivated, the service life
of the defibrillation capacitor
is extended.
MEDUCORE Standard
8 Operator menu
Factory
setting
Deactivated
Activated
100/min
Activated
Deactivated
Activated
2
EN
121