Weinmann MEDUCORE Standard2 Instructions For Use Manual page 55

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

4.3.2 Charging the battery in the device
Requirement
1. Insert battery into the battery compartment.
Specification
External voltage
Battery status
Battery temperature
If the battery is deeply discharged and you charge it in the device,
the battery status indicator will light up red for a short period of
time. It goes out again when the battery status progresses.
Note the methods of storing the battery and the charging
intervals for prolonged storage
page
173).
The expected life of the battery is 2 years. Recommendation:
Replace the battery after 2 years. If battery life has substantially
dropped before then, replace the battery earlier.
If you receive a replacement battery, you need to fully charge
it before the first use.
The portable system is mounted on a wall mounting with
charging interface.
or
The device is connected to the line power via the power supply
unit and charger or via the 12 V network.
When doing so, please note:
Charging starts automatically if the following conditions
are met:
If the device is switched on, the green arrow appears in the
battery status symbol on the display (example:
and the battery status indicator on the device flashes green.
If the device is switched off, only the battery status
indicator flashes green.
The device remains fully ready for use.
(see "13.3 Storing the battery",
Description
At least 11 V
< 95 % charged
Between 0 °C and 45 °C
MEDUCORE Standard
4 Preparation
)
2
EN
55