4 Preparation

4.2 Connecting a power supply

34
EN
MEDUVENT Standard
Failure of therapy or loss of power resulting from use of an
incorrect power supply unit and charger!
If you are using a portable system with the MEDUVENT Standard
and MEDUCORE Standard/MEDUCORE Standard
combination, the devices might lose power when using the 50 W
power supply unit and charger, and the therapy might fail
prematurely.
⇒ Use only the more powerful 100 W power supply unit and
charger (WM 28937) for the MEDUVENT Standard and
MEDUCORE Standard/MEDUCORE Standard
combination.
Failure of therapy due to defective power supply unit and
charger.
A power supply unit and charger which is defective due to shock,
vibration or wet is unable to charge the battery and will thus lead
to failure of the therapy.
⇒ Do not use the power supply unit and charger outdoors.
⇒ Protect the power supply unit and charger from wet.
⇒ Do not use the power supply unit and charger in an emergency
vehicle.
Risk of infection from contaminated power supply unit and
charger
A contaminated power supply unit and charger can lead to
infections.
⇒ Protect the power supply unit and charger from contamination.