9.2 Intervals

Wipe disinfection
Device
Reusable components x
Reusable accessories
Short-term NIBP cuffs

9.3 Hygienic reprocessing of the device

Damage to the device caused by ingress of liquids!
The device is protected from water jets in line with IP55. Ingress of
liquids may damage the device, components and accessories.
⇒ Do not immerse the device, components, or accessories in
liquids.
1. Disconnect the device from the patient.
2. Switch off the device
77).
3. If necessary: Disconnect the device from the line power.
4. Remove the battery.
5. Disconnect the following parts from the device:
Trunk cable with defibrillation electrodes
Pulse oximetry sensor connecting cable with pulse oximetry
sensor
ECG cable
NIBP connecting tube with NIBP cuff
6. Disconnect the defibrillation electrodes from the trunk cable.
9 Hygienic reprocessing
After each use
x
x
x
(see "5.2 Switching the device off", page
MEDUCORE Standard
At least 1x weekly
x
x
x
x
2
EN
145