15 Technical data
Recommended separation distances between portable and mobile
RF telecommunications equipment and MEDUCORE Standard
2
MEDUCORE Standard
variables are controlled. The customer or user of the MEDUCORE Standard
interference by maintaining a minimum distance between portable and mobile RF telecommunication
equipment (transmitters) and the MEDUCORE Standard
maximum output power of the RF telecommunications equipment).
Separation distance according to transmission frequency in m
Rated power of
the RF device in
150 kHz -
W
80 MHz
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
186
EN
MEDUCORE Standard
Delay in treatment due to power supply network faults!
Transient or pulsed conducted interference may cause artifacts in
the ECG signal and thus interfere with device functioning and
delay treatment.
⇒ If there is major interference in the power supply network, only
operate the device with a battery.
Treatment delays due to strong high-frequency interference
signals!
Overly strong high-frequency interference signals can lead to a
false analysis, false measurement results and false alarms and
therefore impair the functioning of the device and delay
treatment.
⇒ Maintain separation distances.
⇒ Maintain a minimum distance of 30 cm.
is intended for use in an electromagnetic environment in which RF interference
150 kHz -
80 MHz in the
ISM bands
0.12
0.38
1.2
3.8
12
2
2
2
can avoid electromagnetic
2
(as recommended below, according to the
When used as a monitor
80 MHz -
800 MHz
0.4
1.3
4
13
40
800 MHz -
2.5 GHz
0.77
2.4
7.7
24
77