15 Technical data

15.6 ECG monitoring system

Specification
Max. patient impedance with 6-lead ECG
Detected heart rate (with ECG derivation via ECG
cable or derivation via defibrillation electrodes)
Suitability for direct use on the heart
Displayed heart rates with pacemaker impulses (with
ECG derivation via ECG cable)
Pacemaker impulses detected by the device:
Amplitude
Pulse width
Overshoot
Device operating mode in the event of a power
interruption lasting > 30 s
Detection of detached electrodes
Noise suppression
Respiration detection
Max. T-wave suppression
Heart rate averaging
Heart rate accuracy (handling irregular cardiac
rhythms)
Inputs and input impedance
Alarm time for tachycardia (as per 60601-2-27)
Response time of heart rate display:
Steep rise
Steep drop
Recovery time after a defibrillation shock
Band width:
6-lead ECG
Pad ECG
Protection against malfunction caused by
electrosurgery
180
EN
MEDUCORE Standard
ECG monitoring system
500 kΩ
30 bpm to 250 bpm
Type CF (6-lead ECG)
If pacemaker impulses are detected, "---" is
displayed instead of the heart rate.
2 mV to 700 mV
0.5 ms to 2 ms
Not detected
The device saves all user settings and restores them
when the device is restarted.
Yes
Yes
No
0.9 mV
The mean heart rate is obtained by averaging up to
7 heart beat intervals, updated with every heart beat
or at least every second
All complexes are detected. The heart rate display is
between the shortest and the longest detected RR
interval.
< 2.5 MΩ for 6-lead ECG
Signal B1: 1 mV 206/min: 9 s
Signal B1: 0.5 mV 206/min: 10 s
Signal B1: 2 mV 206/min: 8 s
Signal B2: 2 mV 195/min: 7 s
Signal B2: 1 mV 195 /min: 7 s
Signal B2: 4 mV 195 /min: 7 s
5 s
7 s
2 s
0.67 Hz to 40 Hz (ECG leads I, II, III, aVR, aVL, aVF)
0.67 Hz to 40 Hz (ECG lead pad)
Yes
2