Risk of injury due to incorrect NIBP cuff!
An incorrectly selected or used NIBP cuff can lead to patient
injuries.
⇒ Attach the NIBP cuff so that the blood supply is not stopped.
⇒ Do not attach the NIBP cuff to a limb with an intravenous
infusion.
⇒ Do not attach the NIBP cuff to a limb with a shunt.
⇒ Do not attach the NIBP cuff to a limb with open wounds or
burns.
⇒ In the case of patients who have undergone a mastectomy, do
not attach the NIBP cuff to the affected side. In the case of
patients who have undergone double mastectomies, attach
the NIBP cuff to the non-dominant arm.
⇒ Do not attach the NIBP cuff to a limb with poor circulation.
2. Select the NIBP cuff which is suitable for the patient's limb as
per the following table.
Designation
Thigh
Thigh
Upper arm
Large Adult plus
Adult plus
Adult
Small Adult
Child
Infant
Color
Limb circumference
Brown
38-50 cm
Dark red
40-55 cm
Dark blue
28-40 cm
Dark blue
23-33 cm
Turquoise
17-25 cm
Green
12-19 cm
Orange
8-13 cm