5 Operation
Result
84
EN
MEDUCORE Standard
Shock not required
The device performs a cardiac rhythm analysis, charges for shock
delivery and outputs the message:
Voice prompt
Stand clear of the patient
Cardiac rhythm is being analyzed
If, based on the cardiac rhythm analysis, the device determines that
a shock is not required, the device outputs the message:
Voice prompt
Shock not recommended
1. Perform cardiopulmonary resuscitation
cardiopulmonary resuscitation", page
The patient does not have a defibrillatable cardiac rhythm. The
device guides you through cardiopulmonary resuscitation by
means of voice prompts, AED instructions and the metronome. It
warns you again to stand clear of the patient after the preset time
has elapsed (120 seconds when delivered), in order to perform
another cardiac rhythm analysis.
Performing cardiopulmonary resuscitation
This section describes cardiopulmonary resuscitation in the AED
mode. Upon delivery, the device performs cardiopulmonary
resuscitation with the following parameters; these, however, can
be adapted by the operator:
Setting
Duration of CPR phase
Ventilation pause
Compression/ventilation
ratio
Start in intubation mode?
CPR voice prompts
Metronome frequency
Start analysis automatically Yes
2
AED instruction
Stand clear of the patient
Analysis
AED instruction
Shock not recommended
(see " Performing
84).
Patient group
for adults
120 s
5 s
30:2
No
No
100/min
Child
120 s
5 s
15:2
No
No
100/min
Yes