2. SPECIFICATIONS AND APPROVALS

2.1. T200H/M OPERATING SPECIFICATIONS

Parameter
Min/Max Range
(Physical Analog Output)
Measurement Units
Zero Noise
Span Noise
Lower Detectable Limit
Zero Drift (24 hours)
Span Drift (24 hours)
Lag Time
Rise/Fall Time
Linearity
Precision
Gas Flow Rates
AC Power
Recorder Offset
Environmental
Analog Outputs
Analog Output Ranges
Analog Output Resolution
Status Outputs
Control Inputs
Alarm outputs
Temperature Range
Humidity Range
07270D DCN7141

Table 2-1: Model T200H/M Basic Unit Specifications

T200H:
Min: 0-5 ppm Full scale
Max: 0-5000 ppm Full scale (selectable,
independent NO, NO
auto ranging supported)
3
ppm, mg/m
(user selectable)
<20 ppb (RMS)
<0.2% of reading (RMS) above 20 ppm
40 ppb (2x noise as per USEPA)
<20 ppb (at constant temperature and voltage.)
<0.5% of reading (at constant temperature and voltage.)
20 s
95% in <60 s (in switching mode)
1% of full scale
0.5% of reading above 5ppm
T200H:
40 cm³/min sample gas through NO
converter & sensor module
3
250 cm
/min ± 10% through bypass
manifold
290 cm³/min total flow ± 10%
O
Sensor option adds 80 cm³/min to total flow through T200H/M when installed.
2
110V-120V, 60 Hz 3.0A
T200H:
220V-240V, 50/60 Hz 3.0A
110V-120V, 60 Hz 3.0A
T200M:
220V-240V, 50/60 Hz 3.0A
± 10%
Installation category (over-voltage category) II; Pollution degree 2
Intended for Indoor Use Only at Altitudes ≤ 2000m
4 user configurable outputs
All Outputs: 0.1 V, 1 V, 5 V or 10 V
Three outputs convertible to 4-20 mA isolated current loop.
All Ranges with 5% under/over-range
1 part in 4096 of selected full-scale voltage (12 bit)
8 Status outputs from opto-isolators, 7 defined, 1 spare
6 Control inputs, 4 defined, 2 spare
2 relay alarms outputs (Optional equipment) with user settable alarm limits
- 1 Form C: SPDT; 3 Amp @ 125 VAC
5 - 40 °C operating range
0-95% RH non-condensing
Specification
T200M:
Min: 0-1 ppm Full scale
Max: 0-200 ppm Full scale (selectable,
, NO
ranges and
independent NO, NO
2
X
auto ranging supported)
T200M:
2
250 cm³/min sample gas through NO
converter & sensor module ± 10%
Rating
175W
155W
55W
75W
, NO
ranges and
2
X
2
Typical Power Consumption
23