Table 3-1: Display Screen And Touch Control Description - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Getting Started
Field
Status
LEDs indicating the states of Sample, Calibration and Fault, as follows:
Name
SAMPLE
CAL
FAULT
Displays the actual concentration of the sample gas currently being measured by the analyzer in the
Conc
currently selected units of measure
Mode
Displays the name of the analyzer's current operating mode
Displays a variety of informational messages such as warning messages, operational data, test function
Param
values and response messages during interactive tasks.
Control Buttons
Displays dynamic, context sensitive labels on each button, which is blank when inactive until applicable.
28

Table 3-1: Display Screen and Touch Control Description

Color
State
Definition
Off
Unit is not operating in sample mode, DAS is disabled.
On
Sample Mode active; Front Panel Display being updated; DAS data
Green
being stored.
Unit is operating in sample mode, front panel display being updated,
Blinking
DAS hold-off mode is ON, DAS disabled
Off
Auto Cal disabled
Yellow
On
Auto Cal enabled
Blinking
Unit is in calibration mode
Off
No warnings exist
Red
Blinking
Warnings exist
Figure 3-3 shows how the front panel display is mapped to the menu charts illustrated in
this manual. The Mode, Param (parameters), and Conc (gas concentration) fields in the
display screen are represented across the top row of each menu chart. The eight touch
control buttons along the bottom of the display screen are represented in the bottom row
of each menu chart.