To change the update rate for an individual analog output channel, press:
DIAG
PREV NEXT
DIAG AIO
SET>
Continue pressing SET> until you reach the
DIAG AIO
<SET SET>
DIAG AIO
DIAG AIO
<SET SET>
Toggle these
DIAG AIO
buttons to set
the data update
0
rate for this
channel.
EXAMPLE
DIAG AIO
0
DIAG AIO
<SET SET> EDIT
Teledyne API - Model T200H/T200M Operation Manual
From the
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
ANALOG I/O CONFIGURATION
ENTR
AOUTS CALIBRATED: NO
CAL
output to be configured
DATA_OUT_2 5V, NXCNC1, NOCAL
EDIT
DATA_OUT_2 OUTPUT: 5V
SET>
EDIT
Continue pressing SET> until ...
DATA_OUT_2 UPDATE: 5 SEC
EDIT
DATA_OUT_2 UPDATE: 5 SEC
0
5
DATA_OUT_2 UPDATE: 30 SEC
3
0
DATA_OUT_2 UPDATE: 30 SEC
EXIT
EXIT
EXIT
EXIT
EXIT
ENTR EXIT
ENTR
EXIT
EXIT