Calibration Procedures
STEP THREE
SAMPLE
RANGE = 50.000 PPM
< TST TST >
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
SETUP X.X
ENTER PASSWORD:818
8
1
186
Activate NO
measurement stability function.
2
CO =X.XXX
CAL
MORE
DIAG ALRM
8
ENTR
Press ENTR first,
then press EXIT 3
times to return to
SAMPLE menu
Teledyne API - Model T200H/T200M Operation Manual
SETUP X.X
SETUP
<PREV
NEXT>
EXIT
SETUP X.X
EXIT
<PREV NEXT> JUMP
SETUP X.X
EXIT
NO
NO2
SETUP X.X
NO
NO2 NOX
0) DAS_HOLD_OFF=15.0 Minutes
JUMP
EDIT PRNT EXIT
Continue pressing NEXT until ...
2) STABIL_GAS=NOX
EDIT
STABIL_GAS:NOX
NOX
O2
STABIL_GAS:NO2
O2
07270D DCN7141
PRNT EXIT
ENTR EXIT
ENTR EXIT