Troubleshooting & Service
7.5.4.3. AC Configuration –Heaters for Option Packages (JP6)
Table 7-5: Power Configuration for Optional AC Heaters (JP6)
JUMPER
COLOR
RED
1
IZS option not available on the T200H/M
242
An O
sensor option includes AC heaters that maintain an optimum operating
2
temperature for key components of those options. Jumper set JP6 is used to connect the
heaters associated with those options to AC power. Since these heaters work with
either 110/155 VAC or 220/240 VAC, there is only one jumper configuration.
HEATER(S)
1
IZS
Permeation Tube
Heater
O2 Sensor Heater
Figure 7-14:
Typical Set Up of AC Heater Jumper Set (JP6)
Teledyne API - Model T200H/T200M Operation Manual
JUMPER
MODEL'S
BETWEEN
1
USED ON
PINS
1 to 8
1
100s, 200s
&
400s
2 to 7
3 to 10
100s & 200s
4 to 9
FUNCTION
Common
Neutral to Load
Common
Neutral to Load
07270D DCN7141