Teledyne T200H User Manual page 67

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API - Model T200H/T200M Operation Manual
3.7.2.2. O
Calibration Method
2
STEP 1 – Set O
span gas concentration :
2
SAMPLE
A1:NXCNC1=100PPM
< TST TST >
CAL
SAMPLE
GAS TO CAL:NOX
NOX
O2
SAMPLE
GAS TO CAL:O2
NOX
O2
The O
span concentration value automatically defaults to
2
If this is not the the concentration of the span gas being
used, toggle these buttons to set the correct concentration
of the O
07270D DCN7141
Set the expected O
span gas concentration.
2
This should be equal to the percent concentration of the O
reporting range (default factory setting = 20.8%; the approximate O
air).
NOX=XXX.X
SETUP
ENTR EXIT
M-P CAL
ENTR
EXIT
M-P CAL
20.8 %.
calibration gases.
2
A1:NXCNC1 =100PPM
<TST TST> ZERO SPAN
CONC
O2 SPAN CONC:20.8%
0
2
0
.8
0
Getting Started
span gas of the selected
2
content of ambient
2
NOX=X.XXX
EXIT
EXIT ignores the new
ENTR EXIT
setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the previous menu.
65

Advertisement

Table of Contents
loading

This manual is also suitable for:

T200m

Table of Contents