Teledyne API - Model T200H/T200M Operation Manual
3.7.2.2. O
Calibration Method
2
STEP 1 – Set O
span gas concentration :
2
SAMPLE
A1:NXCNC1=100PPM
< TST TST >
CAL
SAMPLE
GAS TO CAL:NOX
NOX
O2
SAMPLE
GAS TO CAL:O2
NOX
O2
The O
span concentration value automatically defaults to
2
If this is not the the concentration of the span gas being
used, toggle these buttons to set the correct concentration
of the O
07270D DCN7141
Set the expected O
span gas concentration.
2
This should be equal to the percent concentration of the O
reporting range (default factory setting = 20.8%; the approximate O
air).
NOX=XXX.X
SETUP
ENTR EXIT
M-P CAL
ENTR
EXIT
M-P CAL
20.8 %.
calibration gases.
2
A1:NXCNC1 =100PPM
<TST TST> ZERO SPAN
CONC
O2 SPAN CONC:20.8%
0
2
0
.8
0
Getting Started
span gas of the selected
2
content of ambient
2
NOX=X.XXX
EXIT
EXIT ignores the new
ENTR EXIT
setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the previous menu.
65