Teledyne API - Model T200H/T200M Operation Manual
TC
TERMINATION
TYPE
TYPE
INPUT TC1 (J15)
K
GROUNDED
K
ISOLATED
K
ISOLATED
J
ISOLATED
J
GROUNDED
8.5.7.4. Valve Control
07270D DCN7141

Table 8-5: Typical Thermocouple Settings

OUTPUT
JUMPER
SCALE TYPE
BETWEEN
2 – 12
5mV / °C
4 – 14
2 – 12
5mV / °C
4 – 14
5 – 15
4 – 14
10mV / °C
5 – 15
1 – 11
10mV / °C
3 – 13
5 – 15
1 – 11
10mV / °C
3 – 13
The relay board also hosts two valve driver chips, each of which can drive up four valves.
The main valve assembly in the T200H/M is the NO/NO
assembly mounted right in front of the NO
actuated with 12 V supplied from the relay board and driven by the CPU through the I
bus.
A second set of valves may be installed if the zero/span valve is enabled in the analyzer.
Specialty manifold valves may be present in the analyzer.
PINS
T200H/M with Mini HiCon Converter
T200H/M with Mini HiCon Converter
T200H/M models with Moly Converter
T200H/M models with Moly Converter
T200H/M models with Moly Converter
converter housing. These two valves are
2
Principles of Operation
USED ON
- Auto-zero solenoid valve
X
JUMPER
COLOR
BROWN
GREY
PURPLE
RED
GREEN
2
C
297