Operating Instructions
4.13.7.5. Flow Calibration
Choose between
sample and ozone
flow sensors.
Adjust these values
until the displayed
flow rate equals the
flow rate being
measured by the
independent flow
meter.
162
The flow calibration allows the user to adjust the values of the sample flow rates as they
are displayed on the front panel and reported through COM ports to match the actual
flow rate measured at the sample inlet. This does not change the hardware measurement
of the flow sensors, only the software-calculated values.
To carry out this adjustment, connect an external, sufficiently accurate flow meter to the
sample inlet. Once the flow meter is attached and is measuring actual gas flow, press:
SAMPLE
A1:NXCNC1=100PPM
< TST TST > CAL
SETUP X.X
CFG ACAL DAS RNGE PASS CLK
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
SETUP X.X
ENTER DIAG PASS: 818
8
1
8
DIAG
NEXT
Repeat Pressing NEXT until . .
DIAG
FLOW CALIBRATION
PREV NEXT
DIAG
FLOW SENSOR TO CAL: SAMPLE
SAMPLE OZONE
DIAG FCAL
ACTUAL FLOW: 480 CC / M
0
4
8
Teledyne API - Model T200H/T200M Operation Manual
NOX=XXX.X
SETUP
PRIMARY SETUP MENU
MORE
EXIT
EXIT
ENTR
EXIT
SIGNAL I / O
ENTR EXIT
.
ENTR EXIT
ENTR EXIT
0
ENTR EXIT
Exit at
any time
to return
to main
the
SETUP
menu
Exit returns
to the
previous menu
ENTR accepts the
new value and
returns to the
previous menu
EXIT ignores the
new value and
returns to the
previous menu
07270D DCN7141