Teledyne T200H User Manual page 323

Nitrogen oxides analyzer
Table of Contents

Advertisement

T200H/M and 200EH/EM Menu Trees (05147J DCN7141)
Test Measurement Name
NONOXCONC
not 6
RANGE
not 6
RANGE1
not 6
RANGE2
not 6
RANGE3
STABILITY
2
RESPONSE
SAMPFLOW
OZONEFLOW
PMT
NORMPMT
AUTOZERO
HVPS
RCELLTEMP
BOXTEMP
10
REMBOXTEMP
PMTTEMP
4
MANIFOLDTEMP
8
CO2CELLTEMP
5
O2CELLTEMP
IZSTEMP
CONVTEMP
10
SAMPRESTTEMP
RCELLPRESS
SAMPPRESS
NOXSLOPE
NOXOFFSET
NOSLOPE
NOOFFSET
07270D DCN7141
Table A-3:
Test Measurements
Message Text
Test measurements
3
NO=396.5 NOX=396.5
3
RANGE=500.0 PPB
3
RANGE1=500.0 PPB
3
RANGE2=500.0 PPB
3
RANGE3=500.0 PPB
3
NOX STB=0.0 PPB
5
O2 STB=0.0 PCT
8
CO2 STB=0.0 PCT
RSP=8.81(1.30) SEC
SAMP FLW=460 CC/M
OZONE FL=87 CC/M
PMT=800.0 MV
NORM PMT=793.0 MV
AZERO=1.3 MV
HVPS=650 V
RCELL TEMP=50.8 C
BOX TEMP=28.2 C
REM BOX TMP=30.1 C
PMT TEMP=7.0 C
MF TEMP=50.8 C
CO2 CELL TEMP=50.8 C
O2 CELL TEMP=50.8 C
IZS TEMP=50.8 C
MOLY TEMP=315.0 C
SMP RST TMP=49.8 C
RCEL=7.0 IN-HG-A
SAMP=29.9 IN-HG-A
NOX SLOPE=1.000
NOX OFFS=0.0 MV
NO SLOPE=1.000
NO OFFS=0.0 MV
APPENDIX A-3: Warnings and Test Measurements
Description
Simultaneously displays NO and NOX
concentrations.
D/A range in single or auto-range modes.
D/A #1 range in independent range mode.
D/A #2 range in independent range mode.
D/A #3 range in independent range mode.
Concentration stability (standard deviation
based on setting of STABIL_FREQ and
STABIL_SAMPLES). Select gas with
STABIL_GAS variable.
Instrument response. Length of each
signal processing loop. Time in
parenthesis is standard deviation.
Sample flow rate.
Ozone flow rate.
Raw PMT reading.
PMT reading normalized for temperature,
pressure, auto-zero offset, but not range.
Auto-zero offset.
High voltage power supply output.
Reaction cell temperature.
Internal chassis temperature.
Remote chassis temperature.
PMT temperature.
Bypass or dilution manifold temperature.
CO
sensor cell temperature.
2
O
sensor cell temperature.
2
IZS temperature.
Converter temperature. Converter type is
MOLY, CONV, or O3KL.
Sample restrictor temperature.
Reaction cell pressure.
Sample pressure.
NOX slope for current range, computed
during zero/span calibration.
NOX offset for current range, computed
during zero/span calibration.
NO slope for current range, computed
during zero/span calibration.
NO offset for current range, computed
during zero/span calibration.
A-11