Troubleshooting & Service
7.5.10.3. Status Outputs
SYSTEM_OK
246
The procedure below can be used to test the Status outputs.
1
2
3
Figure 7-15:
Typical Set Up of Status Output Test
1. Connect a cable between the "D" pin and the "" pin on the status output
connector.
2. Connect a 1000 Ω resistor between the "+" pin and the pin for the status output that
is being tested.
3. Connect a voltmeter between the "D" pin and the pin of the output being tested
(Table 7-10).
4. Under the DIAG / SIGNAL I/O menu (Section 4.13.2), scroll through the inputs and
outputs until you get to the output in question. Alternately turn the output on and off.
The Voltmeter will read approximately 5 VDC when the output is OFF.
The Voltmeter will read approximately 0 VDC when the output is ON.
Table 7-10: Status Outputs Pin Assignments
PIN #
1
2
3
4
5
6
7
8
Teledyne API - Model T200H/T200M Operation Manual
V
+ D C
G n d
4
5
6
7
8
D
1000 Ω
STATUS
SYSTEM OK
CONC VALID
HIGH RANGE
ZERO CAL
SPAN CAL
DIAG MODE
LOW
SPARE
+
07270D DCN7141