Operating Instructions
4.13.6.5. Configuring External Analog Inputs (Option) Channels
DIAG
PREV
DIAG AIO
DIAG AIO
DIAG AIO
XIN1 OFFSET:0.00V
< SET
SET>
EDIT
DIAG AIO
XIN1 UNITS:V
< SET
SET>
EDIT
DIAG AIO
XIN1 DISPLAY:OFF
< SET
EDIT
Figure 4-14.
154
To configure the analyzer's external analog inputs option, define for each channel:
•
gain (number of units represented by 1 volt)
•
offset (volts)
•
engineering units to be represented in volts (each press of the touch screen button
scrolls the list of alphanumeric characters from A-Z and 0-9)
•
whether to display the channel in the Test functions
To access and adjust settings for the external Analog Inputs option channels press:
ANALOG I / O CONFIGURATION
NEXT
ENTR
AOUTS CALIBRATED: NO
< SET SET>
CAL
XIN1:1.00,0.00,V,OFF
< SET SET>
EDIT
DIAG AIO
SET>
EXIT
EXIT
EXIT
DIAG – Analog Inputs (Option) Configuration Menu
Teledyne API - Model T200H/T200M Operation Manual
EXIT
Press SET> to scroll to the first
channel. Continue pressing SET>
EXIT
to view each of 8 channels.
EXIT
XIN1 GAIN:1.00V/V
EDIT
EXIT
DIAG AIO
+
0
0
Press to change
Gain value
Press EDIT at any channel
to to change Gain, Offset,
Units and whether to display
the channel in the Test
functions (OFF/ON).
XIN1 GAIN:1.00V/V
1
.0
0
ENTR EXIT
Pressing ENTR records the new setting
and returns to the previous menu.
Pressing EXIT ignores the new setting and
returns to the previous menu.
07270D DCN7141