4.13.4. ANALOG OUTPUTS AND REPORTING RANGES

4.13.4.1. Analog Output Signals Available on the T200H/M

Table 4-17: Analog Output Voltage Ranges with Over-Range Active

The default offset for all ranges is 0 VDC.
07270D DCN7141
The analyzer has four analog output signals, accessible through a connector on the rear
panel.
A1
+
-
0-20 mA current loop
output available for these
channels only
Figure 4-10:
Analog Output Connector Key
The signal levels of each output can be independently configured as follows. An over-
range feature is available that allows each range to be usable from -5% to + 5% of its
nominal scale:
RANGE
MINIMUM OUTPUT
0-0.1 V
0-1 V
0-5 V
0-10 V
Pin assignments for the ANALOG output connector at the rear panel of the instrument:

Table 4-18: Analog Output Pin Assignments

PIN
ANALOG
OUTPUT
1
A1
2
3
A2
4
5
A3
6
7
A4
8
ANALOG OUT
A2
A3
A4
+
-
+
-
+
MAXIMUM OUTPUT
-5 mV
-0.05 V
-0.25 V
-0.5 V
VOLTAGE
CURRENT
SIGNAL
SIGNAL
V Out
I Out +
Ground
I Out -
V Out
I Out +
Ground
I Out -
V Out
I Out +
Ground
I Out -
V Out
Not Available
Ground
Not Available
Operating Instructions
-
+105 mV
+1.05 V
+5.25 V
+10.5 V
129