Teledyne T200H User Manual page 150

Nitrogen oxides analyzer
Table of Contents

Advertisement

Operating Instructions
AIO CONFIGURATION SUBMENU
DIAG
PREV NEXT
DIAG AIO
SET>
Continue pressing SET> until you reach the
DIAG AIO
<SET SET>
DIAG AIO
SET>
Continue pressing SET> until ...
148
To initiate an automatic calibration for an individual output channel, press:
From the
(in the SETUP DIAG menu)
ANALOG I/O CONFIGURATION
ENTR
AOUTS CALIBRATED: NO
CAL
output to be configured
DATA_OUT_2 5V, NXCNC1, NOCAL
EDIT
DATA_OUT_2 RANGE: 5V
EDIT
Teledyne API - Model T200H/T200M Operation Manual
EXIT
EXIT
DIAG AIO
<SET SET>
CAL
DIAG AIO
AUTO CALIBRATING DATA_OUT_2
EXIT
DIAG AIO
<SET SET> CAL
EXIT
DATA_OUT_2 CALIBRATED:NO
DATA_OUT_2 CALIBRATED: YES
07270D DCN7141
EXIT
EXIT