Getting Started
3.5.2. PNEUMATIC CONNECTIONS TO T200H/M BASIC CONFIGURATION
Note
NO
Gas
x
(High Concentration)
MODEL 701
Zero Gas
Generator
Figure 3-13:
44
Figure 3-13 and Figure 3-14 show the most common configurations for gas supply and
exhaust lines to the Model T200H/M analyzer. Please refer to Figure 3-4 for the
locations of pneumatic connections on the rear panel and Table 3-2 for the descriptions.
Sample and calibration gases should only come into contact with PTFE (Teflon)
or glass or materials. They should not come in contact with FEP or stainless
steel materials.
MODEL T700
Gas Dilution
Calibrator
VENT (if no vent
on calibrator)
PUMP
Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator
Teledyne API - Model T200H/T200M Operation Manual
Source of
SAMPLE GAS
Removed during
calibration
SAMPLE
EXHAUST
VENT
here if input
is pressurized
Instrument
Chassis
07270D DCN7141