4.13.7. OTHER DIAG MENU FUNCTIONS

4.13.7.1. Display Sequence Configuration
Note
Table 4-26: T200H/M Available Concentration Display Values
DISPLAY
VALUE
NO
value computed with the slope and offset values for the currently selected
x
NOX
NO
range.
x
NO
value computed with the slope and offset values for NO
x
NXL
1 (Low)
NO
value computed with the slope and offset values for NO
x
NXH
2 (High)
NO value of computed with the slope and offset values for the currently
NO
selected NO range
NO value computed with the slope and offset values for NO reporting range 1
NOL
(Low)
NO value computed with the slope and offset values for NO reporting range 2
NOH
(High)
NO
value of computed with the slope and offset values for the currently
2
N2
selected NO
NO
value computed for with the slope and offset values for NO
2
N2L
range 1 (Low) & N0 reporting range 1 (Low)
NO
value computed for with the slope and offset values for NO
2
N2H
range 2 (High) & NO reporting range 2 (High)
O2
O
concentration value.
2
1
With the following exceptions this will be reporting range 1 (Low) for the appropriate gas type:
• If the analyzer is in calibration mode, this will be the concentration value computed with the slope and offset for which ever range
is being calibrated.
• If the instrument is in either E-Test or O-Test mode, this will be the value computed with the slope and offset values used by
those tests.
2
Only appears if O
sensor option is installed.
2
07270D DCN7141
The model T200H/M analyzer allows the user to choose which gas concentration
measurement and reporting range is to be displayed in the concentration field on the
instrument's front panel display as well as what order and how long each will appear
before the analyzer cycles to the next item on the display list.
This T200H/M is constantly monitoring all of the gas measurements it is
configured to make regardless of which range is being displayed. This feature
merely changes how that display sequence occurs, not how the instrument
makes measurements.
The software permits the user to choose from the following list of display values:
DESCRIPTION
1
1
1
range
2
Operating Instructions
ASSOCIATED DAS
PARAMETER
reporting range
x
reporting range
x
reporting
x
reporting
x
--
NXCNC1
NXCNC2
--
NOCNC1
NOCNC2
--
N2CNC1
N2CNC2
2
O2CONC
155