4.11.7.1. Configuring Ethernet Communication Using DHCP
SETUP X.X
ID
From this point on,
EXIT returns to
COMMUNICATIONS
MENU
SAMPLE
SETUP X.X
SETUP X.X
<SET
SETUP X.X
<SET
SETUP X.X
<SET
SETUP X.X
<SET
SETUP X.X
<SET
SETUP X.X
<SET
110
1. Consult with your network administrator to affirm that your network server is running
DHCP.
2. Access the Ethernet Menu (SETUP>MORE>COMM>INET).
3. After pressing ENTR at the password menu, press SET> to view the DHCP
settings:
COMMUNICATIONS MENU
INET
COM1
COM2
ENTER SETUP PASS : 818
8
1
8
DHCP: ON
SET>
EDIT
INST IP: 0.0.0.0
SET>
GATEWAY IP: 0.0.0.0
SET>
SUBNET MASK: 0.0.0.0
SET>
TCP PORT: 3000
SET>
EDIT
TCP PORT2: 502
SET>
EDIT
HOSTNAME:
EDIT
Teledyne API - Model T200H/T200M Operation Manual
EXIT
ENTR
EXIT
DHCP: ON is
default setting.
If it has been
set to OFF,
press EDIT
EXIT
and set to ON.
EXIT
THE EDIT button is disabled. Each string of octets
should be assigned numbers by the DHCP; if all 0's,
DHCP failed. Consult your network administrator.
EXIT
EXIT
Do not alter unless
directed to by Teledyne
Instruments Customer
Service personnel
EXIT
EXIT
EXIT
SETUP X.X
DHCP: OFF
OFF
ENTR EXIT
SETUP X.X
DHCP: ON
ON
ENTR
EXIT
07270D DCN7141