4.13.5. ANALOG I/O CONFIGURATION

4.13.5.1. The Analog I/O Configuration Submenu.
SUB MENU
AOUTS
CALIBRATED:
DATA_OUT_1:
DATA_OUT_2
DATA_OUT_3
DATA_OUT_4
AIN CALIBRATED
XIN1
.
.
.
XIN8
1
Changes to RANGE or REC_OFS require recalibration of this output.
132
Table 4-21 lists the analog I/O functions that are available in the T200H/M.

Table 4-21: DIAG - Analog I/O Functions

Shows the status of the analog output calibration (YES/NO) and initiates a calibration
of all analog output channels.
Configures the A1 analog output:
1
RANGE
: Selects the signal type (voltage or current loop) and full scale value of the
output.
OVERRANGE: Turns the ± 5% over-range feature ON/OFF for this output channel.
1
REC_OFS
: Sets a voltage offset (not available when RANGE is set to CURRent loop.
1
AUTO_CAL
: Sets the channel for automatic or manual calibration
1
CALIBRATED
: Performs the same calibration as AOUT CALIBRATED, but on this
one channel only.
OUTOUT: Turns the output channel ON/OFF. A signal. Equal to the low end of the
output scale (zero point) is still output by the analyzer, but no data is sent.
DATA: Allows the user to select which DAS parameter to be output.
SCALE: Sets the top end of the reporting range scale for this channel. The analyzer
automatically chooses the units of measure appropriate for the DAS parameter chosen
(e.g. ppm for concentration parameters; in-Hg-A for pressure measurements, etc.)
UPDATE: Sets the time interval at which the analyzer updates the data being output
on the channel.
Same as for DATA_OUT_1 but for analog channel 2 (NO)
Same as for DATA_OUT_1 but for analog channel 3 (NO
Same as for DATA_OUT_1 but for analog channel 4 (O
Shows the calibration status (YES/NO) and initiates a calibration of the analog to digital
converter circuit on the motherboard.
For each of 8 external analog input channels, shows the gain, offset, engineering units,
and whether the channel is to show up as a Test function.
To configure the analyzer's four analog outputs, set the electronic signal type of each
channel and calibrate the outputs. This consists of:
1. Selecting an output type (voltage or current, if an optional current output driver has
been installed) and the signal level that matches the input requirements of the
recording device attached to the channel.
2. Determine if the over-range feature is needed and turn it on or off accordingly.
3. If a Voltage scale is in use, a bipolar recorder offset may be added to the signal if
required (Section 4.13.5).