Teledyne T200H User Manual page 162

Nitrogen oxides analyzer
Table of Contents

Advertisement

Operating Instructions
4.13.7.4. Ozone Generator Override
160
This feature allows the user to manually turn the ozone generator off and on. This can
be done before disconnecting the generator, to prevent ozone from leaking out, or after a
system restart if the user does not want to wait for 30 minutes during warm-up time.
Note that this is one of the two settings in the DIAG menu that is retained after you exit
the menu. (During initial power up TMR (timer) is displayed while the Ozone brick
remains off for the first 30 minutes). Also note that the ozone generator does not turn on
if the ozone flow conditions are out of specification (e.g., if there is no flow through the
system or the pump is broken).
Teledyne API - Model T200H/T200M Operation Manual
07270D DCN7141