Figure 7-13: Typical Set Up Of Ac Heater Jumper Set (Jp2) - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API - Model T200H/T200M Operation Manual
7.5.4.2. AC Configuration – Standard Heaters (JP2)
Table 7-4: Power Configuration for Standard AC Heaters (JP2)
LINE VOLTAGE
110 VAC / 115 VAC
50Hz & 60 Hz
220 VAC / 240 VAC
50Hz & 60 Hz
1
Bypass manifold is built into the reaction cell
07270D DCN7141
Power configuration for the AC the standard heaters is set using Jumper set JP2 (see
Figure 7-4 for the location of JP2).
JUMPER
COLOR
WHITE
BLUE
Figure 7-13:
Typical Set Up of AC Heater Jumper Set (JP2)
HEATER(S)
Reaction Cell / Sample
Chamber Heaters
Mini Hi-Con
Converter
Moly Converter
1
Bypass Manifold
Reaction Cell / Sample
Chamber Heaters
Hi Concentration
Converter
Moly Converter
1
Bypass Manifold
Troubleshooting & Service
JUMPER
BETWEEN
FUNCTION
PINS
1 to 8
Common
2 to 7
Neutral to Load
3 to 10
Common
4 to 9
Neutral to Load
3 to 10
Common
4 to 9
Neutral to Load
5 to 12
Common
6 to 11
Neutral to Load
1 to 7
Load
3 to 9
Load
3 to 9
Load
5 to 11
Load
241