Teledyne API - Model T200H/T200M Operation Manual
Toggle these
buttons to set
ther value of
the desired
07270D DCN7141
To add a zero offset to a specific analog output channel, press:
DIAG
PREV NEXT
DIAG AIO
SET>
Continue pressing SET> until you reach the
DIAG AIO
<SET SET>
DIAG AIO
SET>
DIAG AIO
<SET SET>
DIAG AIO
+
offset.
EXAMPLE
DIAG AIO
DIAG AIO
<SET SET> EDIT
From the
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
ANALOG I/O CONFIGURATION
ENTR
AOUTS CALIBRATED: NO
CAL
output to be configured
DATA_OUT_2 5V, NXCNC1, NOCAL
EDIT
DATA_OUT_2 OUTPUT: 5V
EDIT
Continue pressing SET> until ...
DATA_OUT_2 REC OFS: 0 mV
EDIT
DATA_OUT_2 REC OFS: 0 mV
0
0
0
0
DATA_OUT_2 REC OFS: -10 mV
0
0
1
0
DATA_OUT_2 REC OFS: -10 mV
Operating Instructions
EXIT
EXIT
EXIT
EXIT
EXIT
ENTR EXIT
ENTR
EXIT
EXIT
137