Figure 4-12: Setup For Calibrating Current Outputs - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API - Model T200H/T200M Operation Manual
Note
Figure 4-13:
07270D DCN7141
See Table 3-2 for
pin assignments of
the Analog Out
connector on the
rear panel.
I OUT +
I OUT -
Analyzer
Figure 4-12:
Setup for Calibrating Current Outputs
Do not exceed 60 V between current loop outputs and instrument ground.
If a current meter is not available, an alternative method for calibrating the current loop
outputs is to connect a 250 Ω ±1% resistor across the current loop output. Using a
voltmeter, connected across the resistor, follow the procedure above but adjust the
output to the following values:
V OUT +
V OUT -
ANALYZER
Alternative Setup for Calibrating Current Outputs
mA
Current
Meter
IN
OUT
I IN +
I IN -
Recording
Device
V
Volt
Meter
+DC
Gnd
V IN +
250 O
V IN -
Recording
Device
Operating Instructions
151