Calibration Procedures

5.8. AUTOMATIC CALIBRATION (AUTOCAL)

MODE
DISABLED
ZERO
1
ZERO-LO
1
ZERO-LO-HI
ZERO-HI
1
LO
1
LO-HI
HI
2
O2 –ZERO
2
O2 ZERO-SP
2
O2 SPAN
1
Only applicable if analyzer is equipped with the second span point valve option (52)
2
Only applicable if instrument is equipped wit the O
PARAMETER
TIMER
Turns on the sequence timer
ENABLED
STARTING DATE
Sequence will operate on Starting Date
STARTING TIME
Sequence will operate at Starting Time
DELTA DAYS
Number of days between each sequence trigger. If set to 7, for example, the AutoCal feature
will be enabled once every week on the same day.
DELTA TIME
Incremental delay on each delta day that the sequence starts. If set to 0, the sequence will start
at the same time each day. Delta Time is added to Delta Days for the total time between
cycles.
This parameter prevents the analyzer from being calibrated at the same daytime of each
calibration day and prevents a lack of data for one particular daytime on the days of calibration.
DURATION
Duration of the each sequence step in minutes. This parameter needs to be set such that there
is enough time for the concentration signal to stabilize. The STABIL parameter shows if the
analyzer response is stable at the end of the calibration. This parameter is logged with
calibration values in the DAS.
CALIBRATE
Enable to do a true, dynamic zero or span calibration; disable to do a calibration check only.
RANGE TO CAL
LOW calibrates the low range, HIGH calibrates the high range. Applies only to auto and remote
range modes; this property is not available in single and independent range modes.
198
The AutoCal feature allows unattended, periodic operation of the zero/span valve
options by using the analyzer's internal time of day clock. The AutoCal feature is only
available on the front panel menu (ACAL) if either the zero/span valve or the IZS
option is installed.
AutoCal operates by executing user-defined sequences to initiate the various calibration
modes of the analyzer and to open and close valves appropriately. It is possible to
program and run up to three separate sequences (SEQ1, SEQ2 and SEQ3). Each
sequence can operate in one of three modes or be disabled:
Table 5-2: AutoCal Modes
Disables the sequence
Causes the sequence to perform a zero calibration or check
Causes the sequence to perform a zero calibration or check followed by a mid-span
concentration calibration or check
Causes the sequence to perform a zero calibration or check followed by a mid-span
concentration calibration or check and finally a high-span point calibration or check.
Causes the sequence to perform a zero calibration or check followed by a high-span
point calibration or check.
Causes the sequence to perform a mid-span concentration calibration or check
Causes the sequence to perform a mid-span concentration calibration or check
followed by a high-span point calibration or check
Causes the sequence to perform a high-span point calibration or check.
Causes the sequence to do a zero-point calibration for the O
Causes the sequence to perform a zero calibration of the or check O
by a mid-span concentration calibration or check of the O
Causes the sequence to perform a zero calibration or check of the O
Each mode has seven parameters to control operational details of the sequence:
Table 5-3: AutoCal Attribute Setup Parameters
Teledyne API - Model T200H/T200M Operation Manual
BEHAVIOR
sensor.
2
sensor option (65(.
2
BEHAVIOR
sensor.
2
sensor followed
2
sensor.
2
07270D DCN7141