Teledyne API - Model T200H/T200M Operation Manual
4.11.3.3. COM Port Testing
SAMPLE
A1:NXCNC1=100PPM
< TST TST > CAL
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
SETUP X.X
COMMUNICATIONS MENU
ID INET
COM1
COM2
Select which
COM port to
test.
07270D DCN7141
The serial ports can be tested for correct connection and output in the com menu. This
test sends a string of 256 'w' characters to the selected COM port. While the test is
running, the red LED on the rear panel of the analyzer should flicker.
To initiate the test press the following key sequence.
NOX=XXX.X
SETUP
MORE
EXIT
EXIT
EXIT
SETUP X.X
COM1 MODE:0
SET>
EDIT
SETUP X.X
COM1 BAUD RATE:19200
<SET
SET>
EDIT
SETUP X.X
COM1 : TEST PORT
<SET
TEST
SETUP X.X
TRANSMITTING TO COM1
<SET
TEST
Test runs
automatically
Operating Instructions
EXIT
EXIT
EXIT
EXIT returns to
COMM menu
EXIT
105