Specifications, Agency Approvals, Warranty; Specifications - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
3 SPECIFICATIONS, AGENCY APPROVALS,
WARRANTY

3.1 Specifications

Ranges
Range Modes
1
Noise at Zero
1
Noise at Span
Lower Detectable Limit
3
Zero Drift
Zero Drift
Span Drift
Lag Time
Rise Time
Fall Time
Sample Flow Rate
Linearity
Precision
Temperature Range
Temp Coefficient
Humidity
Voltage Coefficient
Dimensions HxWxD
Weight, Analyzer
Weight, Analyzer
Weight, Ext Pump Pack
Weight, Internal Pump
Power, Analyzer
Power, Analyzer
Power, Ext Pump
Power, Ext Pump
Power, Int Pump
Environmental
Recorder Output
Analog Resolution
Status Option
Measurement Units
1.
As defined by USEPA.
2.
Defined as twice the zero noise level.
3.
At constant temperature and voltage.
4.
Bi-polar.
5.
Electrical rating for CE Mark compliance.
Analyzer Instruction Manual, 02246, Rev. F1
X
In 1ppb increments from 50ppb to 20,000ppb
Single, Independent, AutoRange
0.2 ppb RMS
<0.5% of reading above 50 ppb
2
0.4 ppb RMS
<0.5 ppb/24 hours
1 ppb/7 days
<0.5% FS/7 days
20 sec
95% in <60 sec
95% in <60 sec
500 cc/min. ± 10%
1% of full scale
0.5% of reading
o
5-40
C
< 0.1% per
0-95% RH non-condensing
< 0.1% per V
7" x 17" x 23.6" (18cm x 43cm x 61cm)
43 lbs (20 kg) w/external pump
55 lbs (25 kg) w/internal pump
16 lbs (7 kg)
5 lbs (2 kg)
100 V ~ 50/60 Hz, 120V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 125 watts
5
230 V ~ 50 Hz, 2.5A peak
110 V ~ 60 Hz, 220V~ 50 Hz, 240V~ 50Hz, 295 watts
5
230 V ~ 50 Hz, 2.5A peak
110 v/50/60 Hz, 60 watts
Installation Category (Over-voltage Category) II Pollution Degree 2
4
0-100 mV, 0-1, 5, 10v
1 part in 1024 of selected voltage or current range
12 Status Outputs from opto-isolator
ppb, ppm, ug/m
o
C
3
3
, mg/m
3-1

Advertisement

Table of Contents
loading

Table of Contents