QUESTION
Can I use the DAS system in
place of a strip chart recorder or
data logger?
How often do I need to change
the particulate filter?
How long does the sample pump
last?
Why does my RS-232 serial
connection not work?

7.9. TECHNICAL ASSISTANCE

07270D DCN7141
Yes. Section 4.7 describes the setup and operation of the DAS system in
detail.
Once per week or as needed. Table 6-1 contains a maintenance schedule
listing the most important, regular maintenance tasks. Highly polluted
sample air may require more frequent changes.
The sample pump should last one to two years and the pump head should
be replaced when necessary. Use the RCEL pressure indicator on the front
panel to see if the pump needs replacement.
If this value goes above 10 in-Hg-A, on average, the pump head needs to
be rebuilt.
There are several possible reasons:
• The wrong cable, please use the provided or a generic "straight-
through" cable (do not use a "null-modem" type cable),
• The DCE/DTE switch on the back of the analyzer is not set properly;
make sure that both green and red lights are on,
• The baud rate of the analyzer's COM port does not match that of the
serial port of your computer/data logger. See Section 7.5.12 more
trouble-shooting information.
If this manual and its trouble-shooting / repair sections do not solve your problems,
technical assistance may be obtained from:
Teledyne-API, Technical Support
9970 Carroll Canyon Road, San Diego, CA 92131
Phone: +1 858 657 9800 or 1-800 324 5190
Fax: +1 858 657 9816
Email: sda_techsupport@teledyne.com.
Before you contact Technical Support, fill out the problem report form in Appendix C,
which is also available online for electronic submission at http://www.teledyne-
api.com/forms/.
Troubleshooting & Service
ANSWER