Operating Instructions
4.7.2.2. Editing DAS Data Channels
EXIT will return to the
previous SAMPLE
display.
Moves the
display up &
down the list of
Data Channels
Inserts a new Data
Channel into the list
BEFORE the Channel
currently being displayed
Moves the display
between the
PROPERTIES for this
data channel.
Allows to edit the channel name, see next key sequence.
84
DAS configuration is most conveniently done through the APICOM remote control
program. The following sequence of touchscreen button presses shows how to edit
using the front panel.
SAMPLE
A1:NXCNC1=100PPM
< TST TST > CAL
PRIMARY SETUP MENU
SETUP X.X
CFG
DAS
RNGE PASS CLK MORE
SETUP X.X
VIEW
EDIT
SETUP X.X
8
1
8
Edit Data Channel Menu
SETUP X.X
0) CONC:
PREV NEXT
Deletes The Data
Channel currently
being displayed
SETUP X.X
NAME:CONC
<SET SET> EDIT PRNT
When editing the data channels, the top line of the display indicates some of the
configuration parameters. For example, the display line:
0) CONC : ATIMER, 4, 800
Translates to the following configuration:
Channel No.: 0
NAME: CONC
TRIGGER EVENT: ATIMER
PARAMETERS: Four parameters are included in this channel
EVENT: This channel is set up to record 800 data points.
Teledyne API - Model T200H/T200M Operation Manual
NOX=XXX.X
SETUP
EXIT
DATA ACQUISITION
EXIT
ENTER DAS PASS: 818
ENTR
EXIT
ATIMER,
8,
800
INS
DEL EDIT
PRNT
EXIT
EXIT
Reports the configuration of current
data channels to the RS-232 ports.
Exits to the Main
Data Acquisition
Menu
Exports the
configuration of all
data channels to
RS-232 interface.
EXITS returns to
the previous
Menu
07270D DCN7141