Electrical Connections; Power Connection - Teledyne T200H User Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Getting Started

3.4. ELECTRICAL CONNECTIONS

Note

3.4.1. POWER CONNECTION

ENSURE THAT the voltage and frequency specifications on the rear panel Label of the
instrument ARE compatibLE with the local power before plugging the T200H/M into line
power.
Do not plug in the power cord if the voltage or frequency is incorrect.
32
To maintain compliance with EMC standards, it is required that the cable length
be no greater than 3 meters for all communication connections.
Refer to Figure 3-4 for the location of the rear panel electrical and pneumatic
connections.
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying
at least 10 A current at your AC voltage and that it is equipped with a functioning earth
ground.
WARNING – RISK OF ELECTRIC SHOCK
High voltages are present inside the instrument's chassis.
Power connection must have functioning ground connection.
Do not defeat the ground wire on power plug.
Turn
off
analyzer
subassemblies.
Do not operate with cover off.
Teledyne API - Model T200H/T200M Operation Manual
CAUTION
power
before
disconnecting
or
connecting
electrical
07270D DCN7141